1-5
Section 1 Introduction
Brooks
®
Mf Series
Installation and Operation Manual
X-TMF-Mfi-Mfx-MFC-eng
Part Number: 541B074AAG
August, 2009
Table 1-3 Mf Series Controllers and Meters Terminal Strip Hookup
Certifications:
EMC Directive 89/336/EEC:
Per EN 61326
Hazardous Location Classification
Non-Incendive
Enclosure Type 4X 1/IP66
Ambient Temperature: 5
0
C > Tamb < 65
0
C (T3C) or
5
0
C > Tamb < 55
0
C (T4)
United States and Canada
UL Listed: E73889 Volume 1, Section 17
UL Recognized: E73889 Volume 3, Section 1
Class I, Division 2, Groups A, B, C and D;
Class II, Division 2, Groups F and G;
Suitable for Class III, Division 2
Per UL 1604 and CSA-213
Pressure Equipment Directive (97/23/EC)
See Table 1-1 for further information
TB-1 Label Function Colo
Terminal 1-6 Identification Code**
1 VSUP Supply Voltage Plus (+) See Table 3 Orange
2 SUPCOM Supply Voltage Common Grn/Blk
3 NC Not Used Blue
4 SUPCOM Signal Common Org/Blk
5 VSIG Voltage Signal Output White
6 ISIG Current Signal Output Green
TB-2 Label Function Colo
Terminal 1-9 Identification Code**
1 NC Not Used Blu/Wht
2 GND Chasis Ground Grn/Wht
3 NC Not Used Red
4 VREF Reference Output +5 Vdc Blu/Blk
5 VOR Valve Override Input Blk/Wht*
6 NC Not Used Red/Wht
7 CMDCOM Command Common Black*
8 VCMD Voltage Command Input (Setpoint) Red/Blk*
9 ICMD Current Command Input (Setpoint) Wht/Blk*
TB-3 Label Function Colo
Terminal 1 & 2 Identification Code**
1 NONE Valve Hookup Orange*
2 NONE Valve Hookup Orange*
*These connections used only for controllers (Models Mf50, Mf51 and Mf53)
**Brooks reference